AD3400

The AD3400 is a workstation that is suitable for R&D through to production levels. With its superior positional accuracy and speed, it is foreseeable to begin biochip projects on this system and then increase throughput using the same system. A choice of 3 nests and mechanical shuttle allows users the ability to interchange substrates and process biochips in a semi-automated process.

The proprietary BioJet Plus technology was developed for high speed dispensing. The technology involves (1) the coupling of a high speed micro solenoid valve with a high resolution syringe pump and (2) synchronization of the dispense system with the movements of the stage. The result is an extremely fas dispensing system, which can deliver volumes non-contact from 20 nL to 4 μL in a single dispensed drop.

BioJet Plus allows for flexible biochip development by dispensing to a glass slide, microtiter plate or membrane material.

Features & Benefits
Flexibility
  • Incorporate Multi Dispensing Technologies
  • Aspirate and Dispense Capability
  • Bulk Dispense Capability
Accuracy
  • High Resolution X-Y-Z Positioning
Upgradable
  • Suitable to Add Components for Batch Production Mode
  • Ability to Add Contact and Non-Contact Dispensing Options
  • Ability toAadd Vision Capabilities

Performance
Accuracy of Dispense Volume
  • ± 5% of Target
Precision of Dispense Volume
  • ≤10% CV at 20 nL
Total System Positional Accuracy
  • ±150 µm (typically ±75 µm)
  • SD 50 µm (typically ≤ 25 µm)
Humidity
  • 60 ± 5% RH

Options
  • Up to 16 BioJet Plus Pumps
  • AirJet Dispensing
  • Syringe Dispensing
  • FrontLine Dispensing
  • Pin Dispensing
  • Humidity Control
  • Substrate Nest
    • Glass Slide, Microtiter Plate, or Membrane
  • Vacuum Pump
  • In Line Degasser
  • Ultrasonic Wash Station
  • Barcode Reader
  • Vision System
Specifications
General:
  • Dimensions: 48" x 30" x 39.5" (1291 mm x 762 mm x 1000 mm)
  • Weight: 500 lb (227 kg)
  • Power Requirements: 110/220 VAC; 50/60 Hz
  • Vacuum Requirements: Vacuum Wash Station - 2.1 CFM (~60 CL)
Mechanical:
  • System Precision
    • X, Y and Z-axis are ± 25 µm (although typically < 10 µm)
    • Manual Nest: ± 250 µm
    • Shuttle Nest: ± 25 µm
  • Z-Axis Height
    • Top Plate: ± 127 µm
    • MTP: ± 127 µm
    • Slide: ± 127 µm
  • XY Axis Squareness
    • 90.000 deg ± 0.050 deg
  • Nest to Axis Parallelism
    • X-Axis: ± 127 µm
    • Y-Axis: ± 127 µm
  • Motion Speed
    • X-Axis: 10.0 ± 0.5 mm/sec
    • Y-Axis: 10.0 ± 0.5 mm/sec
    • Z-Axis: 10.0 ± 0.5 mm/sec